Assessment of the Effect of Nano Composite Shield on Radiation Risk Prevention to Breast During Computed Tomography
Assessment of the Effect of Nano Composite Shield on Radiation Risk Prevention to Breast During Computed Tomography
نویسندگان: پریناز محنتی , رضا ملک زاده , بهارک دیوبند , محمد یوسفی سوته
کلمات کلیدی: Nano Bismuth Oxide, Composite Shield, Breast Cancer, Image Quality
نشریه: 16499 , 1 , 17 , 2020
| نویسنده ثبت کننده مقاله |
پریناز محنتی |
| مرحله جاری مقاله |
تایید نهایی |
| دانشکده/مرکز مربوطه |
دانشکده پزشکی |
| کد مقاله |
77517 |
| عنوان فارسی مقاله |
Assessment of the Effect of Nano Composite Shield on Radiation Risk Prevention to Breast During Computed Tomography |
| عنوان لاتین مقاله |
Assessment of the Effect of Nano Composite Shield on Radiation Risk Prevention to Breast During Computed Tomography |
| ناشر |
4 |
| آیا مقاله از طرح تحقیقاتی و یا منتورشیپ استخراج شده است؟ |
خیر |
| عنوان نشریه (خارج از لیست فوق) |
|
| نوع مقاله |
Original Article |
| نحوه ایندکس شدن مقاله |
ایندکس شده سطح یک – ISI - Web of Science |
| آدرس لینک مقاله/ همایش در شبکه اینترنت |
|
| Background: Application of nanomaterial has recently been introduced as a novel method to make a radiation shield to protect
biological damage.
Objectives: The purpose of this study was to evaluate the effectiveness regarding the use of Nano shield on the protection of the
breast in chest CT scan.
MaterialsandMethods: Bismuthoxide nanoparticles were preparedandsynthesized in the Chemistry Laboratory of the University
of Tabriz. Special tests were used to evaluate the properties of nanoparticles. Then, nano-bismuth oxide shields were designed
with the silicon matrix at 10% and 15% ratios in 0.5- and 1-mm thicknesses, respectively. The dosimetry process was performed by
thermoluminescent dosimeters (TLD) dosimeters in the skin and fourth layers of the breast in the chest CT scan. Also, signals and
noises of the images were obtained to assess the effects of shield on image quality. Nano particles size was equal to 24 nm and they
were spherical in shape. The distribution of nanoparticles in the silicone matrix was approved by scanning electron microscope
energy dispersive spectroscopy (SEM-EDS).
Results: Nano bismuth shields reduced the dose value by 9% and 15% in the skin layer related to 10% and 15% bismuth nanoparticles
ratio (0.5 mm thickness), respectively. Also, by increasing shield thickness to 1 mm, the dose of skin layer was reduced by 18% and
24% related to 10% and 15% of bismuth nanoparticles ratio, respectively. Analysis of CT image showed a 10% increase in the noise
related to the 10% bismuth shield.
Conclusion: Thenanobismuthcomposite shield with different ratiosandthicknessesshowsignificant dose reduction in the breast.
This could play a role in breast cancer prevention during chest CT scan. |