Determination of Bisphenol A in Packed Milk and Mineral Water Samples Marketed in Tabriz (Iran) in 2020 Using High-performance Liquid Chromatography-ultraviolet Detector
Determination of Bisphenol A in Packed Milk and Mineral Water Samples Marketed in Tabriz (Iran) in 2020 Using High-performance Liquid Chromatography-ultraviolet Detector
نویسندگان: محبوب نعمتی , محمدرضا افشار مقدم , امین رستم زاده
کلمات کلیدی: Bisphenol A, High-performance liquid chromatography, Water, Milk, Dispersive liquid-liquid microextraction
نشریه: 0 , 4 , 8 , 2021
| نویسنده ثبت کننده مقاله |
محبوب نعمتی |
| مرحله جاری مقاله |
تایید نهایی |
| دانشکده/مرکز مربوطه |
مرکز تحقیقات ایمنی غذا و دارو |
| کد مقاله |
77255 |
| عنوان فارسی مقاله |
Determination of Bisphenol A in Packed Milk and Mineral Water Samples Marketed in Tabriz (Iran) in 2020 Using High-performance Liquid Chromatography-ultraviolet Detector |
| عنوان لاتین مقاله |
Determination of Bisphenol A in Packed Milk and Mineral Water Samples Marketed in Tabriz (Iran) in 2020 Using High-performance Liquid Chromatography-ultraviolet Detector |
| ناشر |
4 |
| آیا مقاله از طرح تحقیقاتی و یا منتورشیپ استخراج شده است؟ |
خیر |
| عنوان نشریه (خارج از لیست فوق) |
Analytical and bioanalytical chemistry Research |
| نوع مقاله |
Original Article |
| نحوه ایندکس شدن مقاله |
ایندکس شده سطح یک – ISI - Web of Science |
| آدرس لینک مقاله/ همایش در شبکه اینترنت |
|
| Bisphenol A (BPA) is a monomer used to produce several plastic resins in food storage containers. Residual BPA in plastics can be
found in filled foods in containers. Because BPA is an endocrine disruptor, its dietary exposure must be estimated. In this method, the
analyte was extracted using a dispersive liquid-liquid microextraction procedure from water and milk samples. For this purpose, a mixture
of acetone and chloroform (142 µl) was dispersed in the water sample or in the aqueous phase obtained from milk after its deproteinization
with trichloroacetic acid. The obtained cloudy solutions were centrifuged in both samples and the sedimentation phase was analyzed by a
high-performance liquid chromatography-ultraviolet detector. Analytical features of the method consist of limits of detection (0.44 ng ml
-1
in water and 0.51 ng ml
-1
in milk samples) and quantification (1.47 ng ml
-1
in water and 1.72 ng ml
-1
in milk samples), linearity
(r2
= 0.992), precision (RSD ≤ 11.1%), and accuracy (RSD ≤ 3%) were studied under final conditions. The results showed that the obtained
results are acceptable. Different milk and water samples were analyzed using the developed method, and BPA was found in several
samples in the ranges of 11.2-32.6 and 9.6-23.5 ng ml
-1 in milk and water samples, respectively. Comparison of the results with the
maximum residue limit established by the European Commission showed that the BPA content exceeded the authorized content. |