Simultaneous homogeneous liquid–liquid microextraction and dispersive liquid–liquid microextraction for extraction of some plasticizers from polymeric containers and aqueous samples
Simultaneous homogeneous liquid–liquid microextraction and dispersive liquid–liquid microextraction for extraction of some plasticizers from polymeric containers and aqueous samples
نویسندگان: محمدرضا افشار مقدم , محبوب نعمتی , علی محبی
کلمات کلیدی: Phthalate esters; di-(2-
ethylhexyl) adipate;
microextraction; aqueous
samples; gas
chromatography; mass
spectrometry
نشریه: 15130 , 0 , 0 , 2020
| نویسنده ثبت کننده مقاله |
محمدرضا افشار مقدم |
| مرحله جاری مقاله |
تایید نهایی |
| دانشکده/مرکز مربوطه |
مرکز تحقیقات ایمنی غذا و دارو |
| کد مقاله |
73946 |
| عنوان فارسی مقاله |
Simultaneous homogeneous liquid–liquid microextraction and dispersive liquid–liquid microextraction for extraction of some plasticizers from polymeric containers and aqueous samples |
| عنوان لاتین مقاله |
Simultaneous homogeneous liquid–liquid microextraction and dispersive liquid–liquid microextraction for extraction of some plasticizers from polymeric containers and aqueous samples |
| ناشر |
5 |
| آیا مقاله از طرح تحقیقاتی و یا منتورشیپ استخراج شده است؟ |
بلی |
| عنوان نشریه (خارج از لیست فوق) |
|
| نوع مقاله |
Original Article |
| نحوه ایندکس شدن مقاله |
ایندکس شده سطح یک – ISI - Web of Science |
| آدرس لینک مقاله/ همایش در شبکه اینترنت |
|
| ABSTRACT
A green mode of simultaneous homogeneous liquid–liquid microextraction combined with dispersive liquid–liquid microextraction assisted
by sodium sulphate has been developed for the extraction and preconcentration of some phthalate esters and di-(2-ethylhexyl) adipate
from different aqueous samples prior to their determination by gas
chromatography-mass spectrometry. In this method, an appropriate
mixture of sodium sulphate solution (phase separation agent and disperser solvent) containing iso-propanol (extraction solvent) is hastily
injected into an aqueous solution containing sodium sulphate and
selected analytes. By doing so, the extractant is dispersed as tiny
droplets into the aqueous phase. After centrifuging extractant containing the analytes is collected on the top of the aqueous phase. Under the
optimum conditions, limits of detection and quantification were
obtained in the ranges of 14–33, and 47–110 ng L–1, respectively. The
enrichment factors and extraction recoveries of the phthalate esters
ranged from 395 to 470 and 79–94%, respectively. |
| نام فایل |
تاریخ درج فایل |
اندازه فایل |
دانلود |
| EDC-int-2.pdf | 1399/07/15 | 958374 | دانلود |