The Effect of Parallel‑hole Collimator Material on Image and Functional Parameters in SPECT Imaging: A SIMIND Monte Carlo Study

The Effect of Parallel‑hole Collimator Material on Image and Functional Parameters in SPECT Imaging: A SIMIND Monte Carlo Study


چاپ صفحه
پژوهان
صفحه نخست سامانه
چکیده مقاله
چکیده مقاله
نویسندگان
نویسندگان
دانلود مقاله
دانلود مقاله
دانشگاه علوم پزشکی تبریز
دانشگاه علوم پزشکی تبریز

نویسندگان: جلیل پیرایش اسلامیان , بابک محمودیان , اسماعیل قره پاپاق

کلمات کلیدی: Collimator material, functional parameters, image quality, Monte Carlo simulation, single‑photon emission computed tomography, simulating medical imaging nuclear detectors

نشریه: 48281 , 3 , 14 , 2015

اطلاعات کلی مقاله
hide/show

نویسنده ثبت کننده مقاله جلیل پیرایش اسلامیان
مرحله جاری مقاله تایید نهایی
دانشکده/مرکز مربوطه دانشکده پزشکی
کد مقاله 71940
عنوان فارسی مقاله The Effect of Parallel‑hole Collimator Material on Image and Functional Parameters in SPECT Imaging: A SIMIND Monte Carlo Study
عنوان لاتین مقاله The Effect of Parallel‑hole Collimator Material on Image and Functional Parameters in SPECT Imaging: A SIMIND Monte Carlo Study
ناشر 4
آیا مقاله از طرح تحقیقاتی و یا منتورشیپ استخراج شده است؟ بلی
عنوان نشریه (خارج از لیست فوق)
نوع مقاله Original Article
نحوه ایندکس شدن مقاله ایندکس شده سطح دو – PubMed
آدرس لینک مقاله/ همایش در شبکه اینترنت

خلاصه مقاله
hide/show

The collimator in single-photon emission computed tomography (SPECT) is a critical component of the imaging system and plays an impressive role in the imaging quality. In this study, the effect of the collimator material on the radioisotopic image and its functional parameters was studied. The simulating medical imaging nuclear detectors (SIMIND) Monte Carlo program was used to simulate a Siemens E.CAM SPECT (Siemens Medical Solutions, Erlangen, Germany) system equipped with a low-energy high-resolution (LEHR) collimator. The simulation and experimental data from the SPECT imaging modality using (99m)Tc were obtained on a point source and Jaszczak phantom. Seventeen high atomic number materials were considered as LEHR collimator materials. In order to determine the effect of the collimator material on the image and functional parameters, the energy resolution, spatial resolution, contrast, and collimator characteristics parameters such as septal penetration and scatter-to-primary ratio were investigated. Energy spectra profiles, full width at half maximums (FWHMs) (mm) of the point spread function (PSF) curves, system sensitivity, and contrast of cold spheres of the Jaszczak phantom for the simulated and experiment systems have acceptability superimposed. The results of FWHM and energy resolution for the 17 collimators showed that the collimator made of 98% lead and 2% antimony could provide the best FWHM and energy resolution, 7.68 mm and 9.87%, respectively. The LEHR collimator with 98% lead and 2% antimony offers the best resolution and contrast when compared to other high atomic number metals and alloys.

نویسندگان
hide/show

نویسنده نفر چندم مقاله
جلیل پیرایش اسلامیاندوم
بابک محمودیانسوم
اسماعیل قره پاپاقچهارم

لینک دانلود مقاله
hide/show

نام فایل تاریخ درج فایل اندازه فایل دانلود
The Effect of Parallel hole Collimator Material on Image and Functional Parameters in SPECT imaging.pdf1399/02/02899683دانلود
project.doc1399/02/0245056دانلود
thesis.doc1399/02/0256832دانلود