Breast Conservation From Radiation Damage by Using Nano Bismuth Shields in Chest Computed Tomography Scan

Breast Conservation From Radiation Damage by Using Nano Bismuth Shields in Chest Computed Tomography Scan


چاپ صفحه
پژوهان
صفحه نخست سامانه
چکیده مقاله
چکیده مقاله
نویسندگان
نویسندگان
دانلود مقاله
دانلود مقاله
دانشگاه علوم پزشکی تبریز
دانشگاه علوم پزشکی تبریز

نویسندگان: پریناز محنتی , محمد یوسفی سوته , رضا ملک زاده , بهارک دیوبند , سهیلا رفاهی

کلمات کلیدی: Keywords: Nano bismuth, Shields, Dose reduction, Chest CT, Image quality

نشریه: 0 , 1 , 6 , 2019

اطلاعات کلی مقاله
hide/show

نویسنده ثبت کننده مقاله پریناز محنتی
مرحله جاری مقاله تایید نهایی
دانشکده/مرکز مربوطه دانشکده پزشکی
کد مقاله 67487
عنوان فارسی مقاله Breast Conservation From Radiation Damage by Using Nano Bismuth Shields in Chest Computed Tomography Scan
عنوان لاتین مقاله Breast Conservation From Radiation Damage by Using Nano Bismuth Shields in Chest Computed Tomography Scan
ناشر 5
آیا مقاله از طرح تحقیقاتی و یا منتورشیپ استخراج شده است؟ خیر
عنوان نشریه (خارج از لیست فوق) CRESCENT JOURNAL OF MEDICAL AND BIOLOGICAL SCIENCES
نوع مقاله Original Article
نحوه ایندکس شدن مقاله ایندکس شده سطح یک – ISI - Web of Science
آدرس لینک مقاله/ همایش در شبکه اینترنت

خلاصه مقاله
hide/show

Abstract Objectives: The present study aimed to calculate the efficiency of bismuth nanoparticles shield to protect the breast in chest computed tomography (CT) scan and investigate the alteration of image quality after using the shield. Materials and Methods: Three groups were included in this experimental research in which phantom was scanned without using nano bismuth shield (group 1), using 0.5 mm nano bismuth shield (group 2), and employing nano bismuth shield with 1 mm thickness (group 3). The polymethyl methacrylate phantom was applied to present the main organ of the chest including the breast and lung. Multi-detector 6-slice CT scan system was utilized for the tests. In addition, the breast radiation dose was measured by ion-chamber which was placed at 12 o’clock position in the phantom according to dosimetry protocol. Further, chest CT image processing was performed once without shield and once in the presence of 10% nano bismuth shield with different thicknesses. Finally, image quality was compared by adding the shield for correct diagnosis. Tests were repeated three time and data were analyzed by ANOVA. Results: Based on the results, using nano bismuth shields, breast dose reduced by 7.2% and 13.8% for 0.5 and 1 mm thickness of the shield, respectively. Furthermore, the CT number and noise increased from 121.3 to 126.36 HU and from 24.53 to 28.1 HU, respectively. Finally, the dose reduction ratio approximately doubled by increasing the thickness from 0.5 to 1 mm. Conclusions: Generally, nano bismuth shields have a noticeable potential for reducing the breast dose with the smallest change in the noise.

نویسندگان
hide/show

نویسنده نفر چندم مقاله
پریناز محنتیاول
محمد یوسفی سوتهدوم
رضا ملک زادهسوم
بهارک دیوبندچهارم
سهیلا رفاهیپنجم

لینک دانلود مقاله
hide/show

نام فایل تاریخ درج فایل اندازه فایل دانلود
yousefi.jpg1398/04/19146572دانلود
2019-Breast Conservation From Radiation Damage by Using Nano.pdf1398/03/29376359دانلود