Impact of Pelvic and Rad-Board Lead Shields on Operator and Patient Radiation Dose in Trans-Radial Coronary Procedures

Impact of Pelvic and Rad-Board Lead Shields on Operator and Patient Radiation Dose in Trans-Radial Coronary Procedures


چاپ صفحه
پژوهان
صفحه نخست سامانه
چکیده مقاله
چکیده مقاله
نویسندگان
نویسندگان
دانلود مقاله
دانلود مقاله
دانشگاه علوم پزشکی تبریز
دانشگاه علوم پزشکی تبریز

نویسندگان: آیدا خالقی فرد , لیلی پورافکاری , مرتضی قوجازاده , علی طریقت نیا , علیرضا فرج الهی

کلمات کلیدی: cardiac catheterization, procedure radiation, dosimetry

نشریه: 55444 , 1 , 187 , 2019

اطلاعات کلی مقاله
hide/show

نویسنده ثبت کننده مقاله علیرضا فرج الهی
مرحله جاری مقاله تایید نهایی
دانشکده/مرکز مربوطه مرکز تحقیقات آموزش علوم پزشکی
کد مقاله 67007
عنوان فارسی مقاله Impact of Pelvic and Rad-Board Lead Shields on Operator and Patient Radiation Dose in Trans-Radial Coronary Procedures
عنوان لاتین مقاله Impact of Pelvic and Rad-Board Lead Shields on Operator and Patient Radiation Dose in Trans-Radial Coronary Procedures
ناشر 6
آیا مقاله از طرح تحقیقاتی و یا منتورشیپ استخراج شده است؟ بلی
عنوان نشریه (خارج از لیست فوق)
نوع مقاله Original Article
نحوه ایندکس شدن مقاله ایندکس شده سطح یک – ISI - Web of Science
آدرس لینک مقاله/ همایش در شبکه اینترنت

خلاصه مقاله
hide/show

Abstract Background Trans-radial approach for cardiac catheterisation procedures has long been associated with high operator and patient radiation dose. The aim of the present study was to determine the effect of pelvic and radial shields on decreasing coronary procedure radiation doses. Methods A total of 418 patients randomly underwent diagnostic and therapeutic cardiac procedures with and without the pelvic and rad-board lead shields during the procedures. The operator and patient doses were then determined by means of a personal dosimeter and dose area product (DAP), respectively. Results The shields decreased the operator radiation dose by 40% in coronary angiography (CA) and by 45% during angioplasty (PCI). These results were achieved at the cost of increased patient radiation dose. Conclusion Pelvic lead shields combined with rad-board shields are highly effective in reducing operator radiation dose in trans-radial approach, but it is only achieved at the cost of increased patient DAP.

نویسندگان
hide/show

نویسنده نفر چندم مقاله
آیدا خالقی فرداول
لیلی پورافکاریسوم
مرتضی قوجازادهچهارم
علی طریقت نیاپنجم
علیرضا فرج الهیششم

لینک دانلود مقاله
hide/show

نام فایل تاریخ درج فایل اندازه فایل دانلود
page.docx1399/11/26136629دانلود
khaleghifard.pdf1399/07/14258665دانلود
123.jpg1399/07/14117453دانلود