Dosimetry of Critical Organs in Maxillofacial Imaging with Cone-beam Computed Tomography

Dosimetry of Critical Organs in Maxillofacial Imaging with Cone-beam Computed Tomography


چاپ صفحه
پژوهان
صفحه نخست سامانه
چکیده مقاله
چکیده مقاله
نویسندگان
نویسندگان
دانلود مقاله
دانلود مقاله
دانشگاه علوم پزشکی تبریز
دانشگاه علوم پزشکی تبریز

نویسندگان: رامین قنبرنژاد فرشی , اصغر مصباحی , معصومه جوهری , ناهیده قره آغاجی

کلمات کلیدی: Cone-Beam Computed Tomography, Radiation Dosimetry, Entrance Surface Dose, Thermoluminescent Dosimetry, Maxillofacial Imaging, Dentistry

نشریه: 19508 , 1 , 9 , 2019

اطلاعات کلی مقاله
hide/show

نویسنده ثبت کننده مقاله ناهیده قره آغاجی
مرحله جاری مقاله تایید نهایی
دانشکده/مرکز مربوطه دانشکده پیراپزشکی
کد مقاله 66812
عنوان فارسی مقاله Dosimetry of Critical Organs in Maxillofacial Imaging with Cone-beam Computed Tomography
عنوان لاتین مقاله Dosimetry of Critical Organs in Maxillofacial Imaging with Cone-beam Computed Tomography
ناشر 5
آیا مقاله از طرح تحقیقاتی و یا منتورشیپ استخراج شده است؟ خیر
عنوان نشریه (خارج از لیست فوق)
نوع مقاله Original Article
نحوه ایندکس شدن مقاله ایندکس شده سطح دو – PMC
آدرس لینک مقاله/ همایش در شبکه اینترنت

خلاصه مقاله
hide/show

Background: While the benefits of cone-beam computed tomography (CBCT) are well known in maxillofacial imaging, the use of this modality is not risk-free. Objective: The aim of this study was to evaluate the exposure doses received by patients during maxillofacial imaging with CBCT. Methods: Entrance surface dose (ESD) was measured by using thermoluminescent dosimeters (TLDs) attached to the eyes lids, parotid glands and thyroid of 64 patients in two imaging centers (A and B). Phantom dosimetry was performed by a cylindrical poly-methyl methacrylate (PMMA) head-size phantom and an ionization chamber for different exposure parameters. NewTom VGi and Planmeca Promax 3D CBCT scanners were used at centers A and B, respectively. Results: The mean ESD of the eyes, parotid glands and thyroid were 2.57, 2.33 and 0.28 mGy in center A, 0.35, 2.11 and 0.37 mGy in center B, respectively. ESD of the eyes revealed a significant difference in two centers; in center B, it was 86.4% lower than center A. In the phantom dosimetry, the measured doses of NewTom VGi were 2.63 and 2.08 mGy, respectively by changing field of view (FOV) size from 8×8 cm2 (height × diameter) to 6×6 cm2. For Planmeca Promax 3D, it ranged from 0.98 to 3.24 mGy depending on exposure parameters. Conclusion: There is a wide range of radiation doses dependent on the units, patients and selected scan parameters. Inappropriate selection of exposure settings, especially FOV size, can seriously increase patient dose.