Assessment of the efficiency of new bismuth composite shields in radiation dose decline to breast during chest CT
Assessment of the efficiency of new bismuth composite shields in radiation dose decline to breast during chest CT
نویسندگان: پریناز محنتی , رضا ملک زاده , محمد یوسفی سوته
کلمات کلیدی: Bismuth-silicon composite shield, Dose reduction, Chest CT, Image quality.
نشریه: 0 , 3 , 5 , 2018
| نویسنده ثبت کننده مقاله |
رضا ملک زاده |
| مرحله جاری مقاله |
تایید نهایی |
| دانشکده/مرکز مربوطه |
تیم های تحقیقاتی دانشگاه |
| کد مقاله |
64606 |
| عنوان فارسی مقاله |
Assessment of the efficiency of new bismuth composite shields in radiation dose decline to breast during chest CT |
| عنوان لاتین مقاله |
Assessment of the efficiency of new bismuth composite shields in radiation dose decline to breast during chest CT |
| ناشر |
4 |
| آیا مقاله از طرح تحقیقاتی و یا منتورشیپ استخراج شده است؟ |
بلی |
| عنوان نشریه (خارج از لیست فوق) |
The Egyptian Journal of Radiology and Nuclear Medicine |
| نوع مقاله |
Original Article |
| نحوه ایندکس شدن مقاله |
ایندکس شده سطح سه – Scopus |
| آدرس لینک مقاله/ همایش در شبکه اینترنت |
|
| Background:Bismuth shield has been recently introduced for radiation protection of patient radiosensitive organs such as breast during chest CT with image diagnosis capability. The purpose of this study was to evaluate
the dose reduction and image quality conserve using new bismuth-silicon composite shields during chest CT.
Materials and methods:Scans were performed on a PMMA phantom using a 6-slice MDCT system. Dose reduction
was calculated by placing ion-chamber in the 12 o’clock position for breast dose measuring. Chest CT scans was
performed with and without new bismuth composite shields 10% by 0.5 mm and 1 mm thicknesses. Image
quality was assessed by CT numbers and noise evaluation.
Results:The results of using bismuth composite shields induced to dose reduction to 7% and 12% for 0.5 mm and
1 mm thicknesses, respectively. As a result of the using 0.5 mm and 1 mm bismuth composite shields, the mean
CT numbers and noise increased by a factor of 0.01 and 0.02 and also from 0.08 to 0.13, respectively. Significant
dose decline was measured and no relevant noise was found.
Conclusion:Results showed that new bismuth-silicon composite shields have good potential to breast dose reduction with smallest noise for diagnosis in chest CT. |