Construction and verification of a physical chest phantom from suitable tissue equivalent materials for computed tomography examinations

Construction and verification of a physical chest phantom from suitable tissue equivalent materials for computed tomography examinations


چاپ صفحه
پژوهان
صفحه نخست سامانه
چکیده مقاله
چکیده مقاله
نویسندگان
نویسندگان
دانلود مقاله
دانلود مقاله
دانشگاه علوم پزشکی تبریز
دانشگاه علوم پزشکی تبریز

نویسندگان: ایمان امینی , پریسا اخلاقی , پروین سربخش

کلمات کلیدی: Physical chest phantom Tissue equivalent materials Hounsfield unit Effective atomic number Electron density

نشریه: 55443 , 1 , 150 , 2018

اطلاعات کلی مقاله
hide/show

نویسنده ثبت کننده مقاله پریسا اخلاقی
مرحله جاری مقاله تایید نهایی
دانشکده/مرکز مربوطه مرکز تحقیقات ایمونولوژی
کد مقاله 62968
عنوان فارسی مقاله Construction and verification of a physical chest phantom from suitable tissue equivalent materials for computed tomography examinations
عنوان لاتین مقاله Construction and verification of a physical chest phantom from suitable tissue equivalent materials for computed tomography examinations
ناشر 3
آیا مقاله از طرح تحقیقاتی و یا منتورشیپ استخراج شده است؟ بلی
عنوان نشریه (خارج از لیست فوق)
نوع مقاله Original Article
نحوه ایندکس شدن مقاله ایندکس شده سطح یک – ISI - Web of Science
آدرس لینک مقاله/ همایش در شبکه اینترنت

خلاصه مقاله
hide/show

Physical phantoms are used to optimize various imaging modalities (e.g. computed tomography) in terms of Xray image quality and absorbed dose. In this regard, this study tried to design and construct a suitable physical phantom with high accuracy and low price in diagnostic energy range. The phantom geometry was selected to be similar to a 5 cm thick cross-sectional slice of an adult human chest. To choose the appropriate tissue-equivalent materials, the physical properties (mass density, electron density and effective atomic number) of a wide range of available polymers were discussed and their accordance with those of human tissues were investigated. Phantom body was made of polymethylmethacrylate and 17 holes were considered in the phantom for placement of different tissue-equivalent materials. According to the physical properties, polypropylene, polyethylene, acrylonitrile butadiene styrene, polyurethane, polyamide and polyoxymethylene were selected and prepared for replacement of adipose, breasts, muscle, liver, cartilage and ribs, respectively. In addition, two different polyurethane foams were made of their raw materials for replacement of the lungs in inhalation and exhalation modes. Then, the prepared phantom was scanned by Siemens somatom sensation 64-slice scanner at tube voltage of 120 kVp and Hounsfield units of tissue-equivalent materials were measured. In addition, using theoretical relationships and goodness of fit test, their Hounsfield units were calculated. It was observed that, the calculated values were able to predict the measured values with the accuracy of 99%. Finally, the Hounsfield units of real human tissues were determined by the mathematical relationship obtained in the previous step; and in order to verify the outcomes, they were compared with other published reports. It was found that the Hounsfield units of real human tissues obtained using this physical phantom and Gammex RMI 465 phantom (scanned by Siemens somatom sensation 64) had the least differences. This demonstrated the high precision of the phantom and selected tissue substitutes, so it could be used in diagnostic energy range.