Comparison of tracheal tube cuff pressure with two techniques: fixed volume versus pilot balloon palpation
Comparison of tracheal tube cuff pressure with two techniques: fixed volume versus pilot balloon palpation
نویسندگان: فرزاد رحمانی , حسن سلیمانپور , عطا محمودپور , کاووس شهسواری نیا , جعفر رحیمی پناهی , سروین سنایی اسکویی , مریم سلیمان پور , رباب مهدیزاده
کلمات کلیدی: Endotracheal Intubation
Tracheal Tube Cuff Pressure
Emergency Department
نشریه: 0 , 9 , 4 , 2017
| نویسنده ثبت کننده مقاله |
حسن سلیمانپور |
| مرحله جاری مقاله |
تایید نهایی |
| دانشکده/مرکز مربوطه |
مرکز مدیریت و پیشگیری از مصدومیتهای حوادث ترافیکی |
| کد مقاله |
62083 |
| عنوان فارسی مقاله |
Comparison of tracheal tube cuff pressure with two techniques: fixed volume versus pilot balloon palpation |
| عنوان لاتین مقاله |
Comparison of tracheal tube cuff pressure with two techniques: fixed volume versus pilot balloon palpation |
| ناشر |
9 |
| آیا مقاله از طرح تحقیقاتی و یا منتورشیپ استخراج شده است؟ |
بلی |
| عنوان نشریه (خارج از لیست فوق) |
Journal of Cardiovascular and Thoracic Research |
| نوع مقاله |
Original Article |
| نحوه ایندکس شدن مقاله |
ایندکس شده سطح یک – ISI - Web of Science |
| آدرس لینک مقاله/ همایش در شبکه اینترنت |
http://jcvtr.tbzmed.ac.ir/PDF/jcvtr-9-169.pdf?t=636508241853208808 |
| Introduction: Filling tracheal tube cuff (TTC) after intubation is necessary to provide a safe airway
in intubated patients. On the other hand, excessive increase or decrease in the pressure of TTC’s
balloon leads into the dangerous complications such as necrosis and/or aspiration. Accordingly,
in the present study, we tried to evaluate the most two common fixed volume and pilot balloon
palpitation methods to control TTC pressure.
Methods: In a prospective cross-sectional study that was carried out in the emergency department
of Tabriz Imam Reza hospital upon 194 patients who needed intubation and from April 2015 to
June 2016. The patients were randomly allocated into two equal groups. For the first the Pilot
Balloon Palpation technique and for the second group 10 cc fixed volume cuff filling technique
was assigned. After that, the pressure was checked with manometer and data were analyzed using
SPSS software.
Results: TTC pressure average in fixed volume group was 44.96±21.77 cmH2O and for palpation
group, it was 118.15±22.15 cmH2O. There was a meaningful difference between two groups in
terms of cuff inside pressure (P value <0.001) and it was meaningfully lower in fixed volume
group than the first one.
Conclusion: The present study showed that pilot balloon palpation or fixed volume method was
not appropriate methods to assess cuff pressure during intubation and the cuff pressure must be
controlled by the manometer. |
| نام فایل |
تاریخ درج فایل |
اندازه فایل |
دانلود |
| jcvtr-9-169.pdf | 1396/10/16 | 401935 | دانلود |